Sample types are used in a sequence to describe each item being processed. You select sample types in the sequence table or as part of a sequence template.
Sample type descriptions:
Sample – A sample with unknown
amounts of analytes being analyzed.
Blank – A sample without any
analytes that is treated like a sample. It is used as a signal-to-noise
reference for all subsequent samples and for system suitability. If
an ISTD method is being used, then the Blank usually contains the
internal standard.
Double
blank – Sometimes
called "solvent blank", it is used to prove that no chromatographic
artifacts are coming from the instrument flow path for that particular
column.
Cal. Std. (Calibration
standard) – A sample with a known amount of analyte that
is used as a reference to create or update a calibration table in
the processing method.
QC check – A sample with a known
amount of analyte that is used to verify and prove that the calibration
is correct.
Spike - A sample
with a known amount of standard added.
Sys Suit - A
sample that is run to demonstrate that system is working correctly
as expected. This sample is typically run before running other samples.