Export OpenLab CDS 2.x raw data as an ASR (Analytical Studio Reviewer) file format.
Prerequisites
To be able to carry out the procedure as described, you need the privilege Processing Method > Edit Post Processing Plugins parameters. Privileges are configured in the Control Panel.
You use LC or GC data.
In the processing method, select the Tools > Post Processing Plugins node.
Choose ASR export from the drop-down list, and click + to add the plugin to the processing method.
Under Custom name, enter a name.
In the method report, the plugin will be shown with this name.
Select the Run plugin check box to run the plugin at the end of processing your data.
Under Peak Match Window (min.), specify a time window (in minutes) to group the peaks of different signals (MS and UV) that are in this time window.
For example, if an injection is opened in the ASR viewer and shows a peak from both an UV signal and a MS signal within the specified time window, then both peaks will be grouped together. If both peaks do not occur within the specified time window, they are not grouped.
Select Output Form to create only one single file containing all injections of a result set. By default, this option is unselected, and a separate ASR file is created for each injection.
Output path: Define a path to save the exported data.
Click
to navigate to the directory.
Click
to select tokens as required. With tokens you create a dynamic path
that will be named, for example, corresponding to the instrument name or result
folder name.
| If you want to export to a network share: Make sure that instrument service user has sufficient privileges to access this folder. For details, see Create Account to Access Network Share in the installation instructions. |
After importing an ASR file, you can find the following parameters under the given location of the ASR viewer:
| Parameters |
ASR Viewer |
|---|---|
| Sample name |
Displayed in the Vials list. |
|
Number of samples | Displayed in the Vials list. |
| Vial number |
Displayed in the Location column of the Vials list. |
| Peak purity |
Displayed in the Asr Summary/Peak Results table:
|
| Sample purity | Prerequisite: Injections are linked to an LC/MS Sample Purity method, the setting Calculate Sample Purity is enabled, and the Base Signal is set to MS. Purity values are displayed in the Purity column of the Purity tab and in the Quality column of the Vial list. For injections with more than one target mass: A target tab (for example, Target 1, Target 2) is displayed for each target mass and displays the sample purity value in the Purity column. |
Other sample parameters are displayed in the Sample tab of the ASR viewer.