If you use time reference compounds, the application corrects the absolute retention time windows based on possible shifts observed by selected time reference compounds.
ShiftRef = ActualRTRef - ExpRTRef
where | |
ShiftRef |
Time shift of the reference compound |
ActualRTRef |
Actual retention time of the reference compound |
ExpRTRef |
Expected retention time of the reference compound |
For compounds that use a time reference, the expected retention time is calculated using an additional factor.
CorrectedExpRT = ExpRT + (ShiftRef * Factor)
where | |
CorrectedExpRT |
Corrected expected retention |
ExpRT |
Expected retention time |
ShiftRef |
Time shift of time reference compound |
Factor |
Factor for compounds with associated time reference compound (Factor) |
For timed groups, the expected start and stop times are calculated accordingly:
Corrected Range Start = Range Start + (ShiftRef * Factor)
Corrected Range End = Range End + (ShiftRef * Factor)
where | |
Corrected Range Start | Corrected start time of timed group |
Range Start | Start time of timed group |
Corrected Range End | Corrected stop time of timed group |
RangeEnd |
Stop time of timed group |
ShiftRef |
Time shift of time reference compound |
Factor |
Factor for timed groups with an associated time reference compound (Factor) |