For a CI Source, you will see a dialog similar to the one shown below. To configure the instrument to use the JetClean function:
JetClean - EI/CI system is automatically selected. The None and JetClean - EI only system are not available.
For Channel A, select the CI reagent gas.
For Channel B, select:
Hydrogen if you are using a JetClean system.
An alternate CI reagent gas if you are not using a JetClean system.
If you selected Hydrogen for Channel B, to allow hydrogen to continue flow after the run is idle, select Turn off hydrogen gas if inactive (min) and set a time for hydrogen flow up to 3000 minutes. This feature is useful when you want to flush with hydrogen over a weekend. This example shows 10 minutes.
If you select Other in either Channel A or B, enter the appropriate gas name in the location provided.
Click OK to return to the 597xMSD Configuration dialog.