Enables the autosampler to draw from a specified Source.
|
Parameter |
Value |
Description |
|---|---|---|
Draw | Default | The default amount specified in the Set Up Agilent Autosampler panel. |
| Maximum | The maximum amount defined by the autosampler configuration. | |
Amount | An amount that overrides the default injection volume. | |
|
from |
Sample | From the position specified in the sample information or sequence/worklist. |
| Location |
From a specified location.The location can be a well or a vial. | |
|
Seat |
From the seat capillary. | |
| Air | From air, above the seat capillary. | |
|
This function is for vial trays only. | ||
| Plug | From plug specified in the Plug Settings | |
| Flushport |
From wash port | |
| Flush | From flush solvent | |
|
Speed | Default | The default speed, specified in the Set Up Agilent Autosampler panel. |
Maximum | The maximum draw speed | |
Speed | A value that overrides the default draw speed. | |
| Rinse | The default rinse speed, specified in Set Up Dual Loop Autosampler | |
|
useOffset | Default | The default draw position, specified in Set Up Dual Loop Autosampler |
Value | The offset of the autosampler needle is calculated from the BOTTOM of the well, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the well bottom with the needle. Use a positive offset to raise the needle higher in the well to clear precipitates on the bottom of the well, or to select one phase of a biphasic sample. The units are millimeters, with 1/10 mm resolution. The range of offsets available is from -2.5 to 35.0 mm. |