Moves to a specified position and ejects a specified amount from the needle.
|
Parameter |
Value |
Description |
|---|---|---|
Eject |
Default | The default amount specified in the Set Up Agilent Autosampler panel. |
| Maximum | The maximum amount defined by the injector configuration. | |
Amount | A specified amount that overrides the default amount. | |
|
to |
Sample | The position specified in the sample information or sequence/worklist. |
| Location |
From a specified location.The location can be a well or a vial. | |
|
This function is for vial trays only. | ||
| Waste | The waste vial. | |
| Flushport |
The flush solvent. | |
| Seat | The seat capillary. | |
| Speed |
Default | The default eject speed specified in the Set Up Agilent Autosampler panel. |
| Maximum | The maximum eject speed. | |
|
Speed | A value that overrides the default eject speed. | |
| Rinse | The default rinse speed, specified in Set Up Dual Loop Autosampler | |
|
useOffset | Default | The default offset position, specified in the Set Up Agilent Autosampler panel. |
Amount | -2.5 to +35 mm below or above the normal offset position |