Moves the autosampler needle as specified.
|
Parameter |
Value |
Description |
|---|---|---|
Needle | Up | Moves the needle to its upmost position. |
|
Seat |
Moves the needle to the seat. | |
|
Into | Moves the needle into a specified Source with a specified Offset. The Offset of the autosampler needle is calculated from the BOTTOM of the well, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the well bottom with the needle. Use a positive offset to raise the needle higher in the well to clear precipitates on the bottom of the well, or to select one phase of a biphasic sample. The unit is in millimeters, with 1/10 mm resolution. The range of offsets available is from -10 to 50 mm. | |
into | Sample | From the position specified in the sample information or sequence/worklist. |
|
Location |
From a specified location.The location can be a well or a vial. | |
|
This function is for vial trays only. | ||
|
Seat | The seat capillary. | |
|
Upmost | The upper position of the needle. This command is equivalent to using the Up parameter. | |
|
useOffset |
Default |
The default draw position, specified in Set Up Dual Loop Autosampler |
Value | The offset of the autosampler needle is calculated from the BOTTOM of the well, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the well bottom with the needle. Use a positive offset to raise the needle higher in the well to clear precipitates on the bottom of the well, or to select one phase of a biphasic sample. The units are millimeters, with 1/10 mm resolution. The range of offsets available is from -10 to 50.0 mm. |