Feed Injection (Online Sample
Manager)
This section is available
only when Feed injection mode has been
selected. The default values depend on the selection in Method
Preset.
Feed
Speed | The
Feed Speed is the speed of delivery of
the sample into the solvent flow.
Adaptive
Select this option to set the Feed Speed to a proportion of the pump flow.
Typically, a high value (default 80%) is used for isocratic analyses; and a low
value (default 5%) is used for gradient
analyses. Fixed Select this
option to set the Feed Speed to a fixed value; enter the value in the
field. |
Flush
Out |
Specifies the volume of solvent to use to flush out the injection
path. Mode Mark the
automatic check box to use the volume of
solvent calculated based on the injection volume and the volumes of the seat
and loop capillaries.
Volume
When the automatic
check box is cleared, specify the flush-out volume in the
field. |