Schedule a tune (LC/MS)
You can schedule either a Checktune or an Autotune to run at a set interval.
The Schedule Tune dialog box has two panes. You select either Checktune
or Autotune in the left pane.
The information in the right pane shows the current schedule for the selected
tune.
The Autotune and Checktune are triggered according to the schedule defined
in the user interface.
- A scheduled Checktune or
Autotune is only run when
a sequence or single sample is not running.
- If the instrument is busy, the scheduled tune is delayed or skipped.
- If the device is offline or in an error state, the scheduled tune
is not run.
- When Autotune and Checktune are scheduled for the same time, Autotune
takes priority, and the scheduled Checktune is ignored.
- If a scheduled Checktune
or Autotune is missed, then
it is made up as soon as possible.
- If multiple Autotunes and Checktunes are missed, only one Autotune
is run when the instrument becomes available.
- If only 1 or more Checktunes are missed, 1 Checktune is run when
the instrument becomes available.
- A message is added to the Activity Log that a scheduled tune
was missed in the time range.
- If Schedule Tune is running, all of the Tune buttons are disabled
except the Abort tune button. For the client/server environment, if
you do not have OpenLab control and a schedule tune is running, all
buttons are disabled.
- The Tune report is generated and opens on each client in the client/server
environment.
- If you try to run a scheduled Autotune, and the current source
is not supported, then a message is logged that states that the Autotune
is not supported with that source.
What do you want to do?
Schedule a Checktune (LC/MS)
Schedule an Autotune (LC/MS)