Identifier
Caffein_percent
Scope
Injection
Type
Double
Description
Compute the ratio of compound "Caffein" by total amount of all other compounds. Caffein is identified on signal VWD1A .
Unit
%
Formula
CurrentInjection.SignalByName("VWD1A").PeakOrGroupByName( "Caffein").Compound_Amount/Sum("Compound_Amount", CurrentInjection.AllIdentifiedPeaks)*100
base-id: 9501856011
id: 9501856011