GC detector settings, ChemStation integrator

GC Detector: TCD

Slope sensitivity

100.000 

Peak width

0.040 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

GC Detector: ADC

Slope sensitivity

20.000 

Peak width

0.040 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

GC Detector: FID

Slope sensitivity

50.000 

Peak width

0.040 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

GC Detector: ECD

Slope sensitivity

100.000 

Peak width

0.080 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

GC Detector: NPD

Slope sensitivity

500.000 

Peak width

0.040 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

GC Detector: FPD

Slope sensitivity

50.000 

Peak width

0.040 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

GC Detector: uECD

Slope sensitivity

500.000 

Peak width

0.080 

Area reject

1.000 

Height reject

1.000 

Shoulders mode

Off

×