The qualifier response ratio (Rsp (%)) is calculated as:
Rsp(Q) | Response of peak in the EIC/SIM of qualifier |
Rsp(Ref) | Response of peak in the reference signal |
The reference signal depends on the parameters set in the processing method (Compounds > Identification > Compound Table tab and
Compounds > Identification > Qualifier Setup tab).
Compound defined on | Value of Qualifier ratio reference signal for compounds on TIC signal | Reference signal* |
|---|---|---|
TIC | MS base peak SIM/EIC | MS base peak SIM/EIC |
TIC | TIC | TIC |
EIC or SIM | Not relevant | EIC or SIM |
* In OpenLab CDS rev. 2.6, for compounds on TIC, the ratio was always calculated on the TIC response.
In rev. 2.7 and 2.8, for compounds on TIC, the ratio was always calculated on the base peak EIC response.