The calibration points are constructed by calculating an amount ratio and a response ratio for each level of a particular compound in the compound table.
The amount ratio is the amount of the compound divided by the amount of the internal standard at this level.
The response ratio is the response (area or height) of the compound divided by the response of the internal standard at this level.
An equation for the curve through the calibration points is calculated using the type of curve model specified in the compound table of the processing method.