The software allows you to quantify an identified compound that is based on a fixed response factor (Manual Factor column). In this case, the compound amount is calculated using the fixed response factor:
Amount = Response * Manual Factor
where | |
Manual Factor |
Fixed response factor |
Response | Response can be Area, Area%, Height, Height%, Rel. Area, or Rel. Amount (see Response type and response factor) |
For details on calculation of concentrations, see Concentration and mass%.
If the compound amount is quantified using the fixed response factor and ISTD, the formula is read as follows:
Relative area = Area / AreaISTD
or:
Relative height = Height / HeightISTD
The amount is then calculated as follows:
Amount = Relative Area * Manual Factor * AmountISTD
or:
Amount = Relative Height * Manual Factor * AmountISTD
For details on calculation of concentrations, see Concentration and mass%.
With RF defined as Response per amount (default setting):
RF = 1 / Manual Factor
With RF defined as Amount per response:
RF = Manual Factor
For more information on the response factor, see Response type and response factor.