Based on the retention time update type (Never, After each run, or After calibration standards) of all identified compounds or timed groups, the expected retention time in the processing method is automatically updated after the corrected expected retention time has been calculated. If the compound can be found based on the corrected value, the corrected value becomes the new expected value in the method.
Retention time updates can be applied with or without time references.
If retention time update is set to After each run or After calibration standards all injections are processed in sequential order. The change in the method will be applied with the next injection and no more parallel processing of non-calibration injections can be done. |
In addition to updating the retention times during the run, you can also manually shift all retention times by a given value.