Symmetry factor, tailing factor (USP)

NOTE

The Symmetry Factor (USP, EP, JP) is identical with the Tailing Factor (USP). Both are available as "Peak_TailFactor" in Intelligent Reporting. See also Pharmacopeia values in OpenLab CDS.

Figure Performance Parameters

S

Symmetry factor, tailing factor (USP)

H

Peak height

tR

Retention time

f

Distance in min between peak front and tR, measured at 5% of the peak height

W50

Peak width at 50% of height [min]

W5

Peak width at 5% of peak height [min]

Wt

Tangent peak width

×