The concentration is shown in the Concentration column in the injection results.
The calculation depends on the following settings (all settings are available in the General tab of the Compounds > Calibration node):
Concentration and corrected amount calculation: with this setting you choose to use dilution factors either as a divisor or as another multiplier.
Calculate mass %: calculate the concentration as a mass percentage (compound amount relative to sample amount).
With Calculate mass % cleared (this is the default setting):
Concentration = Amount * Multipliers * Dilution Factors
or
With Calculate mass % selected:
The concentration is calculated as a mass percentage (compound amount relative to sample amount).
or
where | |
Multipliers |
Sample multpliers or compound multiplier |
Dilution Factors | Sample dilution factors |
Amount |
Calculated amount (including sample-specific compound multipliers) |
For more information on the calculation of multipliers and dilution factors, see Correction factors.