This function enables you to change the peakwidth (response time) at the specified time. The peakwidth is defined as the width of a peak, in minutes, at half the peak height.
You can choose from
same as Peakwidth
2-times Peakwidth (Peakwidth × 2)
4-times Peakwidth (Peakwidth × 4)
8-times Peakwidth (Peakwidth × 8)
When used in the timetable, this parameter changes the filters used for peak-controlled spectral acquisition, but not the data rate of a chromatographic signal.
This setting makes sense only with peak-controlled spectra; it allows you to change the peakwidth setting to account for broadening peaks at the end of the run. |
If the application supports full spectra only (for example, EZChrome) Change Peak Detector Peakwidth is not available. |