Pressure Limits (Integrated LC Isocratic Pump)

Sets the maximum and minimum pressure limits for the pump.

Max

Max is the maximum pressure limit at which the pump will switch itself off, protecting the analytical system against over-pressure.

Min

Min is the minimum limit at which the pump will switch itself off, for example, if any solvent reservoir is empty, this prevents system damage by pumping air.

Limits: You can set the minimum pressure limit to any value between 0 and the upper pressure limit setting.