Select the appropriate volume for the syringe from the drop-down list. You can select from the following:
100 µL, 900 µL
You can enter a value for the seat capillary volume between 0.0 µL and 8000 µL. Agilent provides seat capillaries with the following volumes:
2.3 µL, 400 µL, 1400 µL
If you install a custom seat capillary, enter an appropriate vale for the volume.
The waittime is enabled only in the case of multiple draw, if the seat capillary volume is greater than or equal to the syringe volume. It allows the temporary vacuum, which originates from the drawing of liquid from the sample vial, to dissipate. After injecting sample into the seat capillary, the system waits for the specified time before moving the needle up for the next sample draw.
The maximum injection volume is dependent on the syringe and seat capillary volumes:
If seat capillary volume < syringe volume,then max. injection volume = syringe volume
If seat capillary volume ≥ syringe volume,then max. injection volume = syringe volume + seat capillary volume