Mark the Ignore missing vial check box to specify that, if a vial is missing, the injector ignores it and continues with a 6-second dummy run. The message "Missing vial <x>" is logged, and the system continues with the next injection.
If the Ignore missing vial check box is cleared, the injector generates an error and stops the analysis. The message "Missing vial <x>" is logged, and the sample or sequence is aborted.
Ignore missing vial applies to standard injections only. If a missing vial is encountered in an injector program, an error is generated. |