Modules

The Column Thermostat Cluster panel lists all column thermostat modules that are already specified for the cluster in a five-column table.

Column Compartment Cluster

The Column Compartment Cluster panel lists all column thermostat modules that are already specified for the cluster in the five-column table:

TCC #

The number of the column thermostat module

Product/Serial No.

The product number and serial number of the column thermostat module

Valve Opt.

The type of valve built into the column thermostat module, and its maximum allowed pressure

The following 8-pos/9-port valves are supported:

  • 8-pos/9-port valve 1200 bar (5067-4121)

  • 8-pos/9-port valve 400 bar (5067-4108)

  • 8-pos/9-port bio-inert valve 600 bar (5067-4133)

  • 8-pos/9-port preparative valve 600 bar (5067-4194)

Inlet

Mark this check box to specify the column thermostat module that is the inlet to the cluster. You can specify only one column thermostat module as Inlet.

This check box is disabled if no valve is installed.

If only one high-pressure valve is available, specify the column thermostat module that includes it as the Inlet.

Outlet

Mark this check box to specify the column thermostat module that is the outlet from the cluster. You can specify only one column thermostat module as Outlet.

This check box is disabled if no valve is installed.

Use the and buttons at the right of the table to change the sequence of the submodules in the cluster. Click a selected submodule one place up in the list, or to move it one place down.

Add

Adds a column thermostat to the cluster.

Delete

Deletes the selected column thermostat from the cluster. Deleting all column thermostat modules from the list breaks the cluster.

Configure

Configures the selected column thermostat.