Change Temperature (Column Compartment Cluster Timetable)

Changes in temperature are initiated at the specified time. However, the temperature setpoint takes some time to reach the specified temperature. You cannot specify a temperature gradient for the column compartment.

If you have enabled the As detector cell option in the Temperature section of the method setup, any temperature changes specified in the timetable are ignored.

Limits

-5.0 to 80.0°C (G1316A) or -5.0 to 100.0°C (G1316B/C) in steps of 0.1°C, or Not controlled (temperature control is turned off).

You can enter a more precise value for the temperature, but the temperature is controlled in steps of 1°C to an accuracy of 0.5°C.