The Temperature section allows you to set the temperature of the selected column in the column thermostat cluster. Select the column in the Column section.
Select the Not controlled option to deactivate the temperature control.
Select the middle option to control the temperature; enter the required temperature in the adjacent field.
Limits: -5.0 to 80.0°C (G1316A) or -5.0 to 100.0°C (G1316B/C) in steps of 0.1°C.
The thermostat can only cool down to a temperature 10 ºC below the ambient temperature. |
Select the As Detector Cell option to set the temperature to the cell temperature of the configured detector. This option is available only if a G1315C DAD SL or a G1365C MWD SL is configured. Note that this option is not available for the G1315D DAD or the G1365D MWD. When more than one detector that supports this function is configured, the following priorities are applied:
the G1315C DAD SL has priority over the G1365C MWD SL;
for two detectors of the same type, the one with the lower serial number has priority.