The Valve section allows you to switch the solvent flow between the column and the bypass or a second column for column switching, or to reverse the solvent flow for backflushing.
Port 1 → 6 |
connected ports: 1 with 6, 2 with 3, 4 with 5 |
Port 1 → 2 | connected ports: 1 with 2, 3 with 4, 5 with 6 |
Use current | the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Port 1 → 3 |
connected ports: 1 with 3, 2 with 8, 4 with 6, 5 with 7 |
Port 1 → 7 |
connected ports: 1 with 7, 2 with 4, 3 with 5, 6 with 6 |
Use current | the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Port 1 → 10 |
connected ports: 1 with 10, 2 with 3, 4 with 5, 6 with 7, 8 with 9 |
Port 1 → 2 |
connected ports: 1 with 2, 3 with 4, 5 with 6, 7 with 8, 9 with 10 |
Use current |
the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Position 1 | Inlet port 1 to Outlet port 1' |
Position 2 | Inlet port 2 to Outlet port 2' |
Position 3 |
Inlet port 3 to Outlet port 3' |
Position 4 | Inlet port 4 to Outlet port 4' |
Use current | the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Pos 1, Pos 2, .. Pos 6 | Positions 1 to 6, a separate entry for each position |
Use current | the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Pos 1, Pos 2, .. Pos 8 | Positions 1 to 8, a separate entry for each position |
Use current | the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Port c → n |
center port (c) to pert n (n = 1 to 12) |
Use current | the valve remains in the current position at the end of a run |
Next pos. after run | the valve switches to the next position at the end of a run |
Pos 1, Pos 2, .. Pos n | Positions 1 to 6, a separate entry for each position up to the number of configured positions |
Use current | |
Next pos. after run | the valve switches to the next position at the end of a run |
For the mode selected, the diagram shows how you need to connect your capillaries and columns (except the Generic Valve type). The heavy lines on the diagram indicate the solvent flow path.If a column switching valve is not installed, the Column Switching Valve section is disabled.