Change Right Temperature (Multicolumn Thermostat Timetable)

Enables you to set the temperature of the right zone of the column compartment.

If the Right Temperature setpoint is set to Combined, all timetable entries for Change Right Temperature are ignored.

Changes in temperature are initiated at the specified time. However, the temperature setpoint takes some time to reach the specified temperature. You cannot specify a temperature gradient for the Multicolumn Thermostat.

If you have enabled the As detector cell option in the Temperature section of the method setup, any temperature changes specified in the timetable are ignored.

Limits:

G7116A: 4.0 to 85.0°C in steps of 0.1°C.

G7116B: 0.0 to 110.0°C in steps of 0.1°C.