Valve Position/Column (Multicolumn Thermostat)

Use the Valve Position/Column section to specify the valve position and/or plumbing to be used in the method.

Use Current Column/Position

Choose this option to use the current position of the valve.

Use Selected Column/Position

Choose this position to select a specific position of the valve. Click the down-arrow and select the valve position to use from the table, which shows full details of the occupancy of all valve positions. The selected position is highlighted in the visual representation of the MultiColumn Thermostat.

Click this button to load the plumbing configuration currently stored in the Multicolumn Thermostat. The updated information is displayed in the visual representation of the MultiColumn Thermostat.

Enforce Column for Run

Mark this check box to restrict the execution of the method to a specific column type, identified by its product number. Either enter the product number or select it from the drop down list, which contains all columns that have been defined in the Column Assignment dialog box and contain a product number. Note that product numbers in the Column Assignment dialog box are hidden by default, and can be displayed by clicking >> for the extended table view. Note also that the valve position selected in the Plumbing section of the Column Assignment dialog box still needs to match.

If the Enforce Column for Run check box is marked, the method can be executed only when the system is equipped with the specified column type at the same position as specified during method setup.