Temperature
Settings (Valve Thermostat Cluster)
The
Temperature Settings section
allows you to set the temperature control mode and
Enable Analysis parameters for each
temperature zone in the Valve Thermostat Cluster. The parameters are set out in
a four-column table. The first two columns set the temperature control mode;
the second two columns set the Enable
Analysis parameters.
Zone |
The zone for which you can specify temperature
settings. |
Settings |
Click the down-arrow and select the temperature control
mode from the drop-down list:
Not
Controlled Turns off temperature
control of the zone.
As Detector
Cell Sets the temperature of the
zone to the cell temperature of the configured detector. This option is
available only when a detector with a temperature-controlled cell is
configured. When more than one detector that supports this function is
configured, select the detector that will control the temperature from the
Temperature As Detector Cell section
of the Valve Thermostat Cluster Control.
Combined (Right side
only) Sets the temperature of the
right zone to the same as the left zone.
Controlled
Allows you to set the zone to a specific temperature. Enter the required
temperature in the Temperature
field. Unchanged (MCT and ICC
only) Leaves the temperature
setting unchanged. The zone keeps the setting that is currently stored in the
instrument. |
Temperature
Range | Click the down-arrow and
select the enable analysis mode from the drop-down list:
With any temperature
Select this option to specify that the start of the
analysis is independent of the temperature of the
zone.
When temperature is
within Select this option to
specify that the analysis starts only when the temperature of the zone is
within a specified temperature range. Type the temperature range in the upper
field and, if the column host is a G7116A/B or a G7130A, type an equilibration
time in the lower field. |
When Front Door
Open | Mark the check box in this
column to specify that an analysis can be started irrespective of the state of
the front door of the selected device. When the check box is
cleared, an analysis can be started only when the front door is
closed. |
Enforce
thermal equilibration device (MCT
only) | Mark this check box to
restrict the execution of the method to a Multicolumn Thermostat that is
equipped with a Thermal Equilibration
Device. |