Stoptime (Valve Thermostat Cluster)

Stoptime enables you to set a time at which the module stops an analysis. If the module is used with other Agilent Modular LC modules, the module Stoptime stops the module only and does not stop any other modules.

Limits: 0.01 to 99999.00 minutes or As Pump/Injector (the Stoptime of the pump or injector when an Agilent pump or injector is configured).

When a Stoptime is set for a module and that time is reached during the run, the module stops its run, that is, it stops data acquisition and starts counting the Posttime.

The Stoptime of the whole system is determined as follows:

  • If your system contains an Agilent pump, Agilent valve drive G1170A or Agilent sampler, the Stoptime of the system is determined by the shortest Stoptime set for any of these modules.

  • If your system does not contain any of these modules, the Stoptime is determined by the shortest Stoptime set for any of the remaining Agilent modules in the system.