Change Signal (DAD Timetable)

Enables you to change one or more of the signal settings at the specified time.

Select Signal

Select the signal to change from the drop-down list.

Settings

Wavelength

The wavelength at which the absorbance of the sample is measured as a function of time. The absorbance at the reference wavelength is subtracted from the absorbance at the sample wavelength.

Limits

The precision with which you can set the Wavelength depends on the CDS application.

Bandwidth

The bandwidth of the sample wavelength. The bandwidth determines the wavelength range over which the absorbance is measured. For example, a sample wavelength of 254 nm with a sample bandwidth of 20 nm measures absorbance from 244 to 264 nm.

Limits

The precision with which you can set the Bandwidth depends on the CDS application.

Bandwidth

The bandwidth of the sample wavelength. The bandwidth determines the wavelength range over which the absorbance is measured. For example, a sample wavelength of 254 nm with a sample bandwidth of 20 nm measures absorbance from 244 to 264 nm.

Limits

The precision with which you can set the Bandwidth depends on the CDS application.

Reference Wavelength

The wavelength at which a reference absorbance is measured as a function of time. Mark the check box to switch on the reference wavelength. The reference wavelength compensates for fluctuations caused by changes in baseline absorbance, for example due to changes in solvent composition during gradient elution.

Limits

The precision with which you can set the Reference Wavelength depends on the CDS application.

Reference Bandwidth

The bandwidth of the reference wavelength.

Limits

The precision with which you can set the Reference Bandwidth depends on the CDS application.