Lamp (Modular LC FLD Special Setpoints)

The Lamp section allows you to specify how the lamp is used in an analysis.

Select Only On During Run to specify that the lamp is switched on only during the run time, and is switched off when the run finishes.

Select Economy Mode to run the lamp in a low frequency, low current mode. This increases the lamp lifetime at the expense of sensitivity.

To prolong the life of the lamp, clear the check box if you have a multiple detector configuration and do not want to use the FLD for the next analyses.

If the Lamp on required for analysis check box is marked, but the lamp is off when the run is started, the lamp is switched on automatically.

Lamp Energy Reference

Choose On to switch on a compensation for flash-to-flash variations in lamp intensity using the reference diode. This gives improved signal-to-noise.