Right valve (Modular LC Flexible Cube Control)

Pressure Limit Cluster

Use this section to specify how the pressure limit for the cluster is to be determined.

None

No cluster has been specified.

Use selected

The pressure limit for the cluster is set to that of the selected cluster partner.

Autodetect

The pressure limit for the cluster is set to that of the module with the lowest pressure limit.