Change Contact Control Adjustment (Modular LC Fraction Collector Cluster Timetable)

This function is available for the G1364D Agilent 1260 Infinity Fraction Collector (µS) only.

The Contact Control Adjustment influences the speed of the needle movement from the surface of a MALDI target upwards. This parameter is used for applications using an additional flow (for example, matrix flow) to the main flow.

Increasing the Contact Control Adjustment value (up to 10) increases the needle lift-up speed. Decreasing the value (factor < 1.0) results in a lower needle lift-up speed. The default value is 1.0.