Change Trigger Mode (Modular LC Fraction Collector Cluster Timetable)

You have the following trigger mode options:

Off

Turns off fraction collection. Use this option to turn off fraction collection at the end of the run if you have not specified a Stoptime.

Peak-based

Fractions will be collected based on the peak detection parameters.

For peak-based fraction collection, specify a maximum peak duration (in minutes) in the max. peak duration field. This parameter can be used to stop the fraction collection in cases where the baseline drifts and the signal does not drop below the specified threshold value. Default is 0.5 min; limits are from 0.1 min to 9999.0 min.

Time-based with number of fractions

Specify a number of fractions that you want to collect for each collection period. The time for each fraction is calculated as the Collection Time/#Fractions, where the Collection Time is the time between when the Trigger Mode is set to Time-based and the next mode change or the end of the run.

Time-based with timeslices

Specify a duration (in minutes) for each fraction.