The following fraction modes are available:
Off | Turns off the current fraction collection. Use Off to turn off fraction collection at the end of the run if you have not specified a Stoptime. | |
Time-based, collecting a number of fractions | Fractions are collected between this time and the next change of fraction mode or Off. Specify the number of fractions to collect in the Number of Fractions field. | |
Time-based, collecting time slices | Time-slice fractions are collected between this time and the next change of fraction mode or Off. Specify the duration of the time-slices to collect in the Time slices field. | |
Time-based, collecting volume slices |
Volume-slice fractions are collected between this time and the next change of fraction mode or Off. Specify the volume of the fractions to collect in the Volume slices field. | |
Peak-based |
Fractions are collected based on the peak detection settings. | |
Peak-based, collecting time slices | Time-slice fractions are collected during the elution of a peak, based on the peak detection settings. Specify the duration of the time-slices to collect in the Time slices field. | |
Peak-based, collecting volume slices |
Volume-slice fractions are collected during the elution of a peak, based on the peak detection settings. Specify the volume of the fractions to collect in the Volume slices field. | |
Peak-based with time-slice recovery |
Time-slice fractions are collected between this time and the next change of fraction mode or Off. If the peak detector encounters a peak, the peak is collected independently of the time slices. Specify the duration of the time-slices to collect in the Time slices field. | |
Peak-based with volume-slice recovery | Volume-slice fractions are collected between this time and the next change of fraction mode or Off. If the peak detector encounters a peak, the peak is collected independently of the volume slices. Specify the volume of the fractions to collect in the Volume slices field. | |