Change Trigger Settings (Fraction Collector II Timetable)

The following trigger settings are available in the timetable:

Trigger source

Click the down-arrow and select the trigger source from the drop-down list.

Peak Detection Mode

    Click the down-arrow and select the peak detection mode from the drop-down list. You can select from:

  • Slope

    Detects peaks based on slope values only.

  • Threshold

    Detects peaks based on threshold values only.

  • Threshold and Slope

    Detects peaks based on both threshold and slope values.

Threshold

The height of the peak that triggers collection

Up Slope

The slope of the rising peak. The slope value is based on the first derivative of the signal.

Down Slope

The slope of the falling peak. The slope value is based on the first derivative of the signal.

Upper Threshold

A maximum peak height threshold that is used until the slope values are applied. This ensures that collection continues even for a saturated signal that might be expected to switch collection off.

The value of the Upper Threshold must be greater than that of the Threshold.

Maximum Peak Duration Mode

    Click the down-arrow and select the mode from the drop-down list. You can select from:

  • No Timeout

    The peak duration has no limit.

  • Use Max Peak Duration

    The peak has a maximum duration, set in the Maximum Peak Duration field.

Maximum Peak Duration

The maximum amount of time a peak trigger remains active.