Use the
Trigger Combinations to specify
how multiple peak triggers are combined to start or stop Fraction
Collection.
The following modes are available:
AND
Collection of a fraction is started when
all peak detectors have sent a start trigger, and continues until one detector
sends a stop trigger.
OR
Collection of a fraction is started when at least one peak detector has
sent a start trigger, and continues until all detectors send a stop
trigger.
AND/OR
Collection of a fraction is started
when all peak detectors have sent a start trigger, and continues until all
detectors send a stop trigger (AND condition for
start, OR condition for
stop)
The