Sets the maximum and minimum pressure limits for the pump.
Min is the minimum limit at which the pump will switch itself off, for example, if any solvent reservoir is empty. This prevents system damage caused by pumping air.
Limits: You can set the minimum pressure limit to any value between 0 and the upper pressure limit setting.
The minimum pressure limit is activated only if the pressure starts above the specified value and falls below it. It is not activated if the pressure starts below the specified value. For the system to create an error, the limits must be exceeded for a few seconds. |
Max is the maximum pressure limit at which the pump will switch itself off, protecting the analytical system against over-pressure.
G1361A | 400 bar (5800 psi) | |||
G7161A | 420 bar (6090 psi) | |||
G7161B | Depends on the pump head installed:
|