Generic System (Agilent 1290 Infinity Binary Pump Valve Cluster)

The Generic System section is displayed only when you have selected Generic LC Instrument as the Manufacturer in the Emulation section.

Dwell Volume

Enter a dwell volume for the system in this field.

The Dwell Volume (V) is the product of the gradient delay time (t) and the flow rate (f):

V = t * f

The gradient delay time is the difference in time between the expected gradient and the actual gradient. It is a function of the physical characteristics of the system. See the Agilent 1290 Infinity with ISET User Manual for details of how to measure the dwell volume of your system.