Purge (G4204A only)
The device is purged through the
automatic purge valve (if installed), or through the autosampler (if the
automatic purge valve is not
installed).
Off | Turns off the
purge. |
On |
The device is purged for the specified time (in minutes)
with the specified solvent composition and flow rate.
Time
Enter the purge time (in minutes) in this field.
Limits: 0 to 360.00 min in steps of
0.01. Flow Enter the purge
flow (in mL/min) in this field. Limits: 0.000
to 10.000 mL/min in steps of 0.001.
Composition
Enter the percentages of solvents B, C and D in the fields; solvent A is
calculated as (100 - (%B + %C +
%D)). |