This function allows you to change the flow rate (the rate of movement of eluent along the column) at the specified time. Changes in flow rate occur linearly from either time zero or the time of the last defined change.
Pressure Limit | Flow Limit |
---|---|
1300 bar (18850 psi) |
0.0 to 5.00 ml/min in steps of 0.001 |
Pressure Limit |
Flow Limit |
---|---|
800 bar (11600 psi) |
0.0 to 5.00 ml/min in steps of 0.001 |
Pressure Limit | Flow Limit |
---|---|
1200 bar (17400 psi) * * The pump automatically reduces the pressure if the flow exceeds 2 ml/min |
0.0 to 10.00 mL/min in steps of 0.001 |
Pressure Limit | Flow Limit |
---|---|
upper pressure limit ≤200 bar (≤2900 psi) | 0.0 to 10.00 mL/min in steps of 0.001 |
upper pressure limit >200 bar (>=2900 psi) |
0.0 to 5.00 mL/min in steps of 0.001 |