The 2D-LC Detectors are listed in a three-column table:
Module Identifier |
The type and serial number of the detector. | |||
Dimension |
Click the down-arrow and select the dimension in which the detector is used.
| |||
Transfer volume (µL) | In the first dimension, this is half the volume of the detector flow cell plus the volume between the detector flow cell and the 2D-LC valve. It determines the time between the detection of the peak and the switching of the valve. In the second dimension, this is the volume between the 2D-LC valve and the 2D detector, including the column. |
Displays the Add Detector dialog box, which allows you to add a new detector to the table (for example, a non-CAN detector that was not detected during Autoconfiguration). | |
Removes the selected detector from the list. |