The control dialog box of the Standard pumps defines the seal wash option for the system.
The option is available only when the pump has seal wash capability. The seal wash capability is detected by the module itself.
Off |
no seal wash is used. |
Single wash |
the seal will be purged for a specified time. |
Periodic | a periodic wash will be applied for a defined period in minutes. The ON time in minutes can also be defined; it must be less than or equal to 70% of the period time. |
For the Agilent 1290 Infinity II High Speed Pump (G7120A) and Agilent 1290 Infinity Bio Binary Pump (G7132A, the seal wash is carried out automatically. |