Sets the maximum and minimum pressure limits for the pump.
Max is the maximum pressure limit at which the pump will switch itself off, protecting the analytical system against over-pressure.
G1312A/C Binary Pump |
400 bar (5800 psi) |
G1312B Binary Pump SL | 600 bar (8700 psi) |
G4302A Binary SFC Pump | 600 bar (8700 psi) |
G4782A Binary SFC Pump | 600 bar (8700 psi) |
G4220A Binary Pump | 1200 bar (17400 psi) for flow rates up to 2 mL/min. For flow rates between 2 mL/min and 5 mL/min, the maximum pressure ramps down to 630 bar (9135 psi). |
G4220B Binary Pump VL | 1200 bar (17400 psi) for flow rates up to 1 mL/min. For flow rates between 1 mL/min and 2 mL/min, the maximum pressure ramps down to 1050 bar (15228 psi). |
G7112B Binary Pump | 600 bar (8700 psi) |
G7120A High Speed Pump and G7132A Bio Binary Pump | 1300 bar (18850 psi) for flow rates up to 2 mL/min. For flow rates between 2 mL/min and 5 mL/min, the maximum pressure ramps down to 800 bar (11600 psi). |
Min is the minimum limit at which the pump will switch itself off, for example, if any solvent reservoir is empty. This prevents system damage caused by pumping air.
Limits: You can set the minimum pressure limit to any value between 0 and the upper pressure limit setting.