Purge (G4204A only)

The device is purged through the automatic purge valve (if installed), or through the autosampler (if the automatic purge valve is not installed).

Off

Turns off the purge.

On

The device is purged for the specified time (in minutes) with the specified solvent composition and flow rate.

  • Time

    Enter the purge time (in minutes) in this field.

    Limits: 0 to 360.00 min in steps of 0.01.

  • Flow

    Enter the purge flow (in mL/min) in this field.

    Limits: 0.000 to 10.000 mL/min in steps of 0.001.

  • Composition

    Enter the percentages of solvents B, C and D in the fields; solvent A is calculated as (100 - (%B + %C + %D)).