Use this function to change the peak width (response time) at the specified time. The peak width is defined as the width of a peak, in minutes, at half the peak height.
Limits: When you set the peak width (in minutes), the corresponding response time is set automatically and the appropriate data rate for signal acquisition is selected as shown in the table:
Peak Width (min) | Response Time (sec) | Signal Data Rate (Hz) |
---|---|---|
<0.0063 |
<0.063 | 37 |
>0.0063 | 0.13 | 37 |
>0.013 | 0.25 | 37 |
>0.025 | 0.5 | 18.5 |
>0.05 | 1 | 9.26 |
>0.1 | 2 | 4.63 |
>0.2 | 4 | 2.31 |
>0.4 | 8 |
1.16 |