Sample+ is used for wellplate sample locations only; it defines an offset from the current location. You specify the offsets for Tray, Plate, Row and Column individually. For example, if the current location is P1-A-01, values for Plate, Row and Column of 1 define location P(1+1)-(A+1)-(01+1), which is P2-B-02. The limits are:
Tray |
-1 to +1 |
Only valid if two trays are inserted |
Plate |
-2 to +2 |
Only valid for wellplates |
Row |
-15 to +15 |
|
Column |
-23 to +23 |
|
Note:
Sample+ may not be used with Overlapped Injection.