Ignore Missing Vial | The injector ignores a missing vial and continues with a 6-second dummy run. The message Missing vial <x> is logged, and the system continues with the next injection. |
Abort Current Sequence | The injector generates an error and stops the analysis. The message Missing vial <x>, sequence aborted is logged, and the sample or sequence/worklist is aborted. |
Abort Current Run | The injector generates an error and stops the run. The message Missing vial <x>, run skipped is logged, and the run is skipped. |
The setting applies to standard injections only. If a missing vial is encountered in an injector program, the current Missing Vial setting is ignored, and the whole sequence is aborted. |
Limitations
Not all versions of Chromatography Data Systems (CDS) support Abort Current Run. Selecting this option will trigger an Abort Current Sequence.
Closing the CDS or going offline will save the current Missing Vial setting. But physically turning off the sampler, before closing the CDS or going offline, will not save current changes.