The Draw Speed determines the rate at which the plunger draws sample from the vial. Set the speed to an appropriate value for your sample. For viscous samples, use a slow Draw Speed.
Type |
Syringe (μL) |
Min (μL/min) |
Max (μL/min) |
---|---|---|---|
Standard autosampler |
100 |
10 |
1000 |
Standard autosampler |
900 |
90 |
1000 |
|
|
|
|
μ autosampler |
8 |
0.7 |
20 |
μ autosampler |
40 |
0.7 |
250 |
The Eject Speed determines the rate at which the plunger ejects sample from the metering device. If you are injecting large volumes of sample, setting a high Eject Speed will shorten the time needed for an injection cycle. For viscous samples, use a slow Eject Speed.
See Draw Speed above
The Draw Position is a vertical offset that enables you to position the needle a specific distance (in mm) away from its standard position. The Draw Position offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample.
The default position (Draw Position = 0.0) is a vertical offset of +2.0 mm. |
Standard autosampler |
-2.5 to 35 mm relative to the standard position |
μ autosampler |
-2.5 to 35 mm relative to the standard position |