Auxiliary (Standard Autosampler)

Draw Speed

The Draw Speed determines the rate at which the plunger draws sample from the vial. Set the speed to an appropriate value for your sample. For viscous samples, use a slow Draw Speed.

Limits:

Eject Speed

The Eject Speed determines the rate at which the plunger ejects sample from the metering device. If you are injecting large volumes of sample, setting a high Eject Speed will shorten the time needed for an injection cycle. For viscous samples, use a slow Eject Speed.

Limits:

See Draw Speed above

Draw Position

The Draw Position is a vertical offset that enables you to position the needle a specific distance (in mm) away from its standard position. The Draw Position offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample.

The default position (Draw Position = 0.0) is a vertical offset of +2.0 mm.

Limits: