The Thermostat section is enabled only when the Thermostat controlled by method check box in the Options section of the Configuration dialog box is marked.
Select the upper option and enter a temperature in the field to specify a sample temperature. You can type a temperature directly into the field, or use the up and down arrows to select a value. When you select this option, the When temperature is within +/- 1ºC option in the Enable Analysis section is automatically selected, and the analysis can be started only when the temperature of the sample is within 1°C of the specified value. This temperature setting overrides that of the autosampler configuration.
You can set the temperature between 4°C and 40°C in steps of 1°C.
The thermostat can cool down to a temperature 10°C below ambient only. |
Select this option to turn off temperature control.